Kumar, Rajesh and Kumar, Harish and Kumar, Anil and -, Vikram (2012) Long Term Uncertainty Investigations of 1 MN Force Calibration Machine at NPL, India (NPLI). Measurement Science Review, 12 (4). pp. 149-152. ISSN 1335-8871

[img]
Preview
PDF - Published Version
Download (161Kb) | Preview

Abstract

The present paper is an attempt to study the long term uncertainty of 1 MN hydraulic multiplication system (HMS) force calibration machine (FCM) at the National Physical Laboratory, India (NPLI), which is used for calibration of the force measuring instruments in the range of 100 kN - 1 MN. The 1 MN HMS FCM was installed at NPLI in 1993 and was built on the principle of hydraulic amplifications of dead weights. The best measurement capability (BMC) of the machine is +/- 0.025% (k = 2) and it is traceable to national standards by means of precision force transfer standards (FTS). The present study discusses the uncertainty variations of the 1 MN HMS FCM over the years and describes the other parameters in detail, too. The 1 MN HMS FCM was calibrated in the years 2004, 2006, 2007, 2008, 2009 and 2010 and the results have been reported.

Item Type: Article
Additional Information: Copyright for this article belongs to M/s Versita.
Subjects: Instruments/ Instrumentation
Divisions: UNSPECIFIED
Depositing User: Users 27 not found.
Date Deposited: 13 Mar 2020 11:27
Last Modified: 13 Mar 2020 11:27
URI: http://npl.csircentral.net/id/eprint/3510

Actions (login required)

View Item View Item