Rani, A. and Upadhyay, R. S. and Singh, Y. P. (2013) Investigating Temperature Distribution of Two Different Types of Blackbody Sources Using Infrared Pyrometry Techniques. Mapan, 28 (2). pp. 91-98. ISSN 0970-3950

[img] PDF - Published Version
Restricted to Registered users only

Download (597Kb) | Request a copy

Abstract

A comparison between two different types of thermal radiation sources maintaining near blackbody conditions has been carried out in the range from 50 to 500 A degrees C. An infrared total radiation pyrometer was used as a transfer standard to measure the temperature of blackbodies. A thorough study of temperature distribution has been carried out for the large surface source in order to characterize the best location over the surface blackbody for temperature determination precisely of the order of better than 0.1 A degrees C. The expanded uncertainty in the estimation of temperature of the radiating source in the above range of measurement was evaluated to be within +/- 0.24 A degrees C at 50 A degrees C and +/- 0.88 A degrees C at 500 A degrees C. The blackbody temperature sources found to be suitable for calibration of infrared total radiation pyrometers and thermal imaging devices in the operational range as mentioned above for laboratory use or other industrial and medical applications.

Item Type: Article
Additional Information: Copyright for this article belongs to M/s Springer Verlag.
Subjects: Instruments/ Instrumentation
Applied Physics/Condensed Matter
Divisions: UNSPECIFIED
Depositing User: Users 27 not found.
Date Deposited: 22 Jul 2020 09:17
Last Modified: 22 Jul 2020 09:17
URI: http://npl.csircentral.net/id/eprint/3157

Actions (login required)