-, Satish and Sawhney, N. and Kumar, S. and Saxena, A. K. (2015) Evaluation of Three Terminal Capacitance Standards at CSIR-NPL. Mapan, 30 (4). 261-265 . ISSN 0970-3950

[img] PDF - Published Version
Restricted to Registered users only

Download (401Kb) | Request a copy

Abstract

The reported work is the part of the establishment of metrological traceability of high frequency impedance standards at CSIR-NPL. The evaluation of three terminal GR 1404 type air capacitance standards of nominal values 100 and 1000 pF has been completed using electrical equivalent circuit model. This involves the measurement of reference capacitance at 1 kHz and residual parameters which are capacitive and inductive in nature. Frequency dependent inductance is estimated using linear curve fit from measured residual inductive parameters. Then frequency dependent capacitance value for capacitance standards has been computed at set of frequencies ranging from 1 kHz to 1 MHz.

Item Type: Article
Uncontrolled Keywords: Reference capacitance; Residual parameter s; Linear curve fit; Frequency dependent inductance; Capacitance standard
Subjects: Instruments/ Instrumentation
Applied Physics/Condensed Matter
Divisions: UNSPECIFIED
Depositing User: Dr. Rajpal Walke
Date Deposited: 23 Sep 2016 05:13
Last Modified: 23 Sep 2016 05:13
URI: http://npl.csircentral.net/id/eprint/1814

Actions (login required)

View Item View Item