-, Babita and Narula, L. K. and John, Thomas and Saleem, M. and Saxena, A. K. (2014) Preliminary Study for the Establishment of DC Charge Calibration Facility at CSIR-NPL Using Charged Capacitance Source. Mapan, 29 (4). 269-272 . ISSN 0970-3950

[img] PDF - Published Version
Restricted to Registered users only

Download (314Kb) | Request a copy

Abstract

The present paper reports the details of the preliminary study towards the establishment of DC charge calibration facility at CSIR-NPL down to pico-coulomb ranges using charged capacitance source. The work involves the development of an automated electric charge generation and measurement setup using automation measurement program developed in LabVIEW programming environment. Charge measurements down to 100 pC have been performed with very good measurement accuracy and the expanded uncertainty has been evaluated as per ISO GUM document.

Item Type: Article
Additional Information: Copyright for this article belongs to M/S Springer Verlag.
Uncontrolled Keywords: Automation Charge measurement Charge amplifier Electrometer Calibration Uncertainty
Subjects: Instruments/ Instrumentation
Physics
Divisions: UNSPECIFIED
Depositing User: Dr. Rajpal Walke
Date Deposited: 03 Nov 2015 06:10
Last Modified: 03 Nov 2015 06:10
URI: http://npl.csircentral.net/id/eprint/1473