-, Babita and -, Satish and Singh, Ajeet and Saxena, A. K. (2013) Determination and Validation of Average Value of National Standard of DC Voltage Using Different Methods at CSIR-NPLI. Mapan, 28 (2). 99-104. ISSN 0970-3950
|
PDF
- Published Version
Restricted to Registered users only Download (410Kb) | Request a copy |
Abstract
The aim of this paper is to evaluate and validate the average value of national or secondary standard of DC voltage, using four different methods; comparison, substitution, difference and ratio method. In the paper the custom made automation program developed to control the whole measurement process is discussed as well. The automation system has been developed in the view of achieving maximal quality of metrological performance i.e. minimum measurement uncertainty and maximum reliability. It is found that measurement automation provides reliable and efficient way for the calibration of Zener based standard of DC voltage