Kumar, Anil and Kumar, Harish (2011) Stability Studies of Torque Transducers. Measurement Science Review, 11 (2). pp. 41-44. ISSN 1335-8871

[img]
Preview
PDF - Published Version
Download (87Kb) | Preview

Abstract

The present paper discusses the stability studies carried out on a small number of torque transducers over the years with capacities ranging from 10 Nm to 1000 Nm. The torque transducers have been calibrated using the torque standard machine based on the written standard calibration procedure. The study reveals that the uncertainty of measurement of torque transducers has been varying appreciably and it is more for lower range. Besides, the deviation from their average values has also been studied and found to be varying through the years.

Item Type: Article
Uncontrolled Keywords: torque transducer; torque standard machine; stability
Subjects: Instruments/ Instrumentation
Divisions: UNSPECIFIED
Depositing User: Mr. Abhishek Yadav
Date Deposited: 29 Oct 2012 08:59
Last Modified: 29 Oct 2012 08:59
URI: http://npl.csircentral.net/id/eprint/768

Actions (login required)

View Item View Item