Saha, Shibu and Arya, Sunil K. and Singh, S. P. and Sreenivas, K. and Malhotra, B. D. and Gupta, Vinay (2009) Nanoporous cerium oxide thin film for glucose biosensor. Biosensors and Bioelectronics, 24 (7). pp. 2040-2045. ISSN 0956-5663
PDF
- Published Version
Restricted to Registered users only Download (1384Kb) | Request a copy |
Abstract
Nanoporous cerium oxide (CeO2) thin film deposited onto platinum (Pt) coated glass plate using pulsed laser deposition (PLD) has been utilized for immobilization of glucose oxidase (GOx). Atomic force microscopy studies reveal the formation of nanoporous surface morphology of CeO2 thin film. Response studies carried out using differential pulsed voltammetry (DPV) and optical measurements show that the GOx/CeO2/Pt bio-electrode shows linearity in the range of 25–300 mg/dl of glucose concentration. The low value of Michaelis-Menten constant (1.01 mM) indicates enhanced enzyme affinity of GOx to glucose. The observed results show promising application of the nanoporous CeO2 thin film for glucose sensing application without any surface functionalization or mediator.
Item Type: | Article |
---|---|
Additional Information: | Copyright for this article belongs to M/s Elsevier B.V. |
Uncontrolled Keywords: | Glucose; Cerium oxide; Biosensor |