Yadav , Sanjay and Bandyopadhyay, A. K. (2009) Evaluation of laboratory performance through interlaboratory comparison. MAPAN - Journal of Metrology Society of India, 24 (2). pp. 125-138. ISSN 0974-9853
|
PDF
- Published Version
Download (9Mb) | Preview |
Abstract
This paper reports the performance of the calibration results obtained by 21 laboratories using digital pressure calibrator as an artifact in the pressure range 7-70 MPa. National Physical Laboratory (NPLI), New Delhi has coordinated this programme and also acted as a reference laboratory. The program started in May, 2006 and completed during May, 2008. The comparison was carried out at 10 equally spaced pressure points i.e. 7, 14, 21, 28, 35, 42, 49, 56, 63 and 70 MPa throughout the entire pressure range of 7 - 70 MPa. The calibration results thus obtained were analyzed as per ISO/IEC GUM document. The 92.7% measurement results are found in agreement with the results of NPLI. The relative deviations between laboratories values and reference values are well within the 0.05% for 123 measurement points, 0.1% for 162 measurement points and 0.25% for 177 measurement points. The difference of the laboratories values with reference values are found almost well within the uncertainty band of the reference values at 68.0% measurement results, within their reported expanded uncertainty band at 81.5% measurement results and within the combined expanded measurement uncertainty band at 92.7% measurement results. Overall, the results are considered to be reasonably good being the first proficiency testing for most of the participating laboratories.
Item Type: | Article |
---|---|
Additional Information: | Copyright for this article belongs to M/s Springer Verlag. |
Subjects: | Instruments/ Instrumentation Physics |
Divisions: | UNSPECIFIED |
Depositing User: | Ms Neetu Chandra |
Date Deposited: | 28 Aug 2012 08:34 |
Last Modified: | 28 Aug 2012 08:34 |
URI: | http://npl.csircentral.net/id/eprint/497 |
Actions (login required)
View Item |