Yadav, Sanjay and Prakash, Om and Gupta, V. K. and Bandyopadhyay, A. K. (2006) Studies on the stabilities of various types of industrial pressure measuring devices. Journal of Scientific and Industrial Research, 65 (9). 721-724 . ISSN 0022-4456

[img]
Preview
PDF - Published Version
Download (155Kb) | Preview

Abstract

Present paper describes studies carriled out on an industrial dead weight tester (pressure range, up to 330 MPa; calibration; 3 times: period, 6 years), one pressure dial gauge (pressure range, up to 70 MPa; calibration, 3 time period, 3 years), and one digital pressure calibrator (pressure range, up to 200 MPa; calibration, 4 times; period, 6 years). A small change was observed in the short-term stability but the change was prominent in case of long-term stability in all the instruments. Stability of dead weight tester and pressure dial gauge was well within their claimed manfacturer Uncertain ties while it was found to be beyond claimed uncertainty in case of digital pressure calibrator.