Singh, J. and Sharma, N. Dilawar and Kumar, A. and Bandyopadhyay, A. K. (2014) Report of the Proficiency Testing in the Pneumatic Pressure Region up to 5 MPa. Mapan, 29 (3). 213-222 . ISSN 0970-3950

[img] PDF - Published Version
Restricted to Registered users only

Download (976Kb) | Request a copy

Abstract

We report the performance of calibration laboratories in the pneumatic pressure region (0 to 5) MPa through a recently conducted inter-laboratory comparison. Six NABL (National Accreditation Board for Testing and Calibration Laboratories) accredited laboratories participated in the program. The proficiency testing program was organized and piloted by CSIR-National Physical Laboratory. The program started in June 2010 and was completed in October 2011. The artifact used was a high precision pressure dial gauge. The reference values were generated by the pilot laboratory. The deviations for each participating laboratory were estimated against the reference pressure values and the compatibility of the results was calculated using the conventional method. Out of the total measurements made, all but two were found to be in good agreement with the reference values. The normalized error values (E (n)) of five laboratories out of the total six were found well within +/- A 1 over the entire pressure range.

Item Type: Article
Additional Information: Copyright for this article belongs to M/S Springer Verlag.
Uncontrolled Keywords: Proficiency testing Pneumatic pressure Quality control
Subjects: Instruments/ Instrumentation
Divisions: UNSPECIFIED
Depositing User: Dr. Rajpal Walke
Date Deposited: 04 Nov 2015 07:25
Last Modified: 04 Nov 2015 07:25
URI: http://npl.csircentral.net/id/eprint/1497

Actions (login required)