-, Rajesh and Sharma, Vikash and Tanwa, Vinod Kumar and Mishra, S. K. and Biradar, A. M. (2010) Electrochemical impedance immunosensor for the detection of cardiac biomarker Myogobin (Mb) in aqueous solution. Thin Solid Films, 519 (3). pp. 1167-1170. ISSN 0040-6090
PDF
- Published Version
Restricted to Registered users only Download (593Kb) | Request a copy |
Abstract
A label-free, electrochemical impedance immunosensor based on surface modified thin flat gold wire electrode is reported for the quantitative detection of cardiac biomarker Myoglobin in aqueous solution. The protein antibody, ab-Mb, was covalently immobilized through a self assembled monolayer of 11-mercaptoundecanoic acid (MUA) and 3-mercapto propionic acid (MPA) via carbodiimide coupling reaction using N-(3-dimethylaminopropyl)-N′-ethyl carbodiimide hydrochloride (EDC) and N-Hydroxy Succinamide (NHS). The immunosensor (ab-Mb/MUA-MPA/Au) was characterized by electrochemical techniques. The electrochemical performance of the immunosensor was studied by electrochemical impedance spectroscopy. The immunosensor showed an increased electrontransfer resistance on coupling with biomarker protein antigen, ag-Mb, in the presence of a redox probe [Fe (CN)6]3−/4−. The modified Au electrode immunosensor exhibits an electrochemical impedance response to antigen, ag-Mb concentrations in a linear range from 10 ng to 650 ng mL−1 with a lowest detection limit of 5.2 ng mL−1.
Item Type: | Article |
---|---|
Additional Information: | Copyright for this article belongs to M/s Elsevier B.V. |
Uncontrolled Keywords: | Immunosensor; Cyclic voltammetry; Electrochemical impedance spectroscopy; Covalent immobilization; Immunoassay |
Subjects: | Materials Science Physics |
Divisions: | UNSPECIFIED |
Depositing User: | Ms Neetu Chandra |
Date Deposited: | 12 Apr 2012 07:10 |
Last Modified: | 12 Apr 2012 07:10 |
URI: | http://npl.csircentral.net/id/eprint/120 |
Actions (login required)
View Item |